Multichannel-systems Roboocyte2 Manual Manual do Utilizador Página 106

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 115
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 105
Roboocyte2 Manual
102
Result Display: Dose-Response Curve
Click on the "Try to fit a Logistic Curve on the Data Points" icon
to open the dose-response
"Logistic Fit (4P)" section of the “Result Display”.
Start the curve fitting by a click on the button
. The start parameters for the fit are
calculated automatically. If the fit does not converge, you can try to type in parameters manually.
After the successful fitting the fit parameters, "Lower Asymptote", "Upper Asymptote"
"EC50/IC59", and "Slope/Steepness" are displayed.